Previous revision
namespace:running_the_laser [2026/04/16 17:52] – [Inlite] adminnamespace:running_the_laser [2026/04/23 15:47] (current) – [Running the Experiment] admin
Line 18: Line 18:
  
 in Mainrunner, I set tele_type='beamprofiler', which leads to teledyne_1 being connected, but I wanted teledyne_3. So I created, tele_type='OAB' which leads to teledyne_3 being connected. in Mainrunner, I set tele_type='beamprofiler', which leads to teledyne_1 being connected, but I wanted teledyne_3. So I created, tele_type='OAB' which leads to teledyne_3 being connected.
 +
 +
 +if I want to run my N shots from the mainrunner script. I need to set Inhibit to Trig, Flip switch to T0, State to Frequency. and (in mainrunner) whattodog='manual'.
 +
 +if you want to use the the Oscilloscope, The RUN button much be Red, when you connect to it. It will not connect properly when the SINGLE button in lit Green.
 +
 +in the Mirror box. there are USB ports in the back. They are known to be glitchy, so much that , devices won't be recognized. If you have, a camera that's plugged into it, and try to connect, it say "No device available..."
 +
 +
 +Here are the 2 commands to show and move the ThorLab Motors for the Nozzle. "trans" can be changed to "foc" and "vert'
 +  * tm_trans.send_new_position(NUMBER)
 +  * tm_trans.
 +
 +
 +====Mazzler, Dazzler, Wizzler====
 +
 +
 +